Ranges:
Standard:
40 nT, 400 nT, 4
:
T, 40
:
T,
400
:
T full scale
*
Alternate Scale:
0.4 mG, 4 mG, 40 mG, 400
mG, 4 G full scale
*
With A/m option:
3 0 m A / m e t e r , 3 0 0
mA/meter, 3 A/meter, 30
A/meter, 300 A/meter full
scale
*
*
Multiply all values by 100 when using the optional X100
probe
Accuracy:
± 5% at calibration
frequency
Linearity:
± 2% Full Scale
Recorder Out Level:
0 - 2.5 VDC, 25
:
A
maximum
Recorder Out Jack:
¼ inch stereo phone jack
Battery:
12 VDC, 1400 mA-h
rechargeable Nickel-Cadmium
(NiCd)
Battery Life:
Approximately 20 Hrs.
continuous operation
(full charge)
Battery Charger:
110/220 VAC, 16 hour
Environmental Operating
Temperature:
0°C to 50°C (+32°F to
+122°F)
Humidity:
5% to 95% relative humidity,
noncondensing