ETS-Lindgren HI-3702 Induced Current Meter User Manual
Page 30

30
|
Series H-491198-36 Battery Charger
Environmental
Operating Temperature:
10
°
C–40
°
C (50
°
F–104
°
F)
Humidity:
5% to 95% relative humidity,
non-condensing
M
AINTENANCE
R
ECOMMENDATIONS
•
Operate the battery charger with care.
•
There are no user serviceable parts inside the battery charger.
Opening the battery charger housing may void your warranty.
R
EPLACING THE
F
USE
Disconnect the battery charger from power
before replacing a fuse.
If the battery charger fails to operate, check for a blown fuse inside the power
entry module. A blown fuse must be replaced with the same value and type of
fuse, or an unsafe condition may result. Use only 250 Volt, 1.0 Amp, Type T
(5 mm x 20 mm) fuses.
To replace a fuse:
1.
Two fuses are located in the fuse drawer in the power input module.
Use a screwdriver to open the drawer.
2.
The fuse towards the outside of the drawer is the spare. Remove the
spare fuse from the module.