7. APPENDIX B: HI-4455 OPERATING

PROTOCOLS

The information in this appendix assumes that you have purchased the
optional HI-4413P Fiber Optic/RS232 Interface, and are capable of
communicating directly with the HI-4455 probe via computer. No system
readout is required when using this configuration.

COMMUNICATION PROTOCOL

Data Type:

RS-232 Serial

Data Mode:

Asynchronous

Word Length:

7 bit

Parity:

Odd

Stop Bits:

1

Data Rate:

9600 baud

INFORMATION TRANSFER PROTOCOL

The HI-4455 operates as a Controller Mode device. This probe only
responds to commands from another device; it transmits no data without
receiving instructions to do so.

COMMAND STRUCTURE

A command to an HI-4455 probe consists of

1) a command letter, followed by
2) possible

parameters

3) terminated with a carriage return.