7. Product Description

Medi Client IIA - Instructions for use (Version 1.03)

7.7.2.1.

DC Power Cable Connector (Option)

The Medi Client IIA can be ordered with a DC power plug terminal (3-pin Phoenix connector).
For the DC connection prepare the connecting wires using the Phoenix plug terminal (optional part).

2

3

4

6

5

1

Fig. 29: Phoenix plug terminal (option) with “plus” and “minus” marking

1 3-pin Phoenix plug terminal

2 Cover over the slotted pan head screws

3 Marking “minus”

4 Marking “plus”

5 Location for inserting the “minus” wire

6 Location for inserting the “plus” wire

1.

Cut two isolated wires to the required length [AWG18 (∅ minimum 1 mm

2

; maximum length: 5 m].

2.

Strip each end 5 –7 mm.

3.

Twist the striped wire-ends and tin it with solder.

4.

Open the cover to have access to the slotted pan head screws.

5.

Loosen the two slotted pan head screws (that correspond to the marked location “+” and “-“ of the DC plug terminal)
far enough so that you can insert the end of the prepared wires.

6.

Insert the wires into the corresponding clamp of the Phoenix plug terminal. Make sure that you have the right polarity
of the connection (refer to Fig. 29).

7.

Fasten the screws to secure the wires into the clamps of the plug terminal.

8.

Close the cover.

The second end of each wire will be prepared as required for the connection to the 24 DC power source.

The DC power source to be used with the Medi Client IIA system must be galvanically isolated according
to IEC60601-1 safety of medical devices!

If you are not sure, use the AC/DC adapter provided with the system.