8 test mode – Rice Lake Intrinsically Safe User Manual

Page 61

Advertising
background image

320IS Installation Manual - Appendix A

55

7.8

Test Mode

In addition to normal and setup modes, test mode provides a number of diagnostic functions for the

320IS

,

including:

•

Display raw A/D count

•

Calibrate A/D offset and gain

•

Display digital input states

•

Reset configuration parameters to default values

•

Transmit test character (“U”) from serial port

•

Display characters received by external serial port

•

Set analog output state to zero or full scale

To enter test mode, press and hold the setup switch until the front panel display shows the word

TEST

. After about

three seconds, the test mode display automatically shifts to the first test menu function, A/DTST.
Figure 7-4 shows the Test Menu structure; Figure 7-5 shows the front panel key functions in test mode. Note that,

because the Test Menu functions are all on a single menu level, the

GROSS/NET

( ) key has no function. Press the

ZERO

( ) key to exit test mode.

Table 7-9 on page 56 summarizes the test menu functions.

Figure 7-4. Test Menu

Figure 7-5. Front Panel Key Functions in Test Mode

A/DTST

TEST

ADCAL-

ADCAL+

D/I 1

D/I 2

D/I 4

D/O 1

D/O 2

D/I 3

D/O 3

D/O 4

A/O 1

A/O 2

DEFLT

XMTEDP

XMTPRN

ECHO R

Exit

Not Used

Enter

Move Left

Move Right