6 correcting the internal mass value – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 26

Advertising
background image

24

7-6 Correcting the Internal Mass Value

The MC series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.

The internal mass value is corrected as follows:

Model

Correction reference value

Correction range

MC-1000

500.0000 g

±0.020 g

MC-6100

2000.000 g

±0.20 g

Example: Using the MC-6100









Operation

1 Calibrate the balance using the internal mass (one-touch

calibration). Then, place an external weight and confirm
the value to be corrected.

In the example, the value is off by -0.060 g and the

correction reference value of the MC-6100 is 2 kg. So, the
value is to be corrected by 0.030 g in 2 kg.


2 Press the ON:OFF key to turn off the display.

3 While pressing and holding the PRINT key and the

SAMPLE key, press the ON:OFF key. The balance

displays p5 .









Correct the internal mass

by +0.030 g in 2 kg.

Calibrate the balance with

the corrected internal mass.

4 kg

4 kg

Use the same weight

With these keys

held down

Press the key