MSI-4300 Porta-Weigh Plus/7 Key
•
User Guide
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
Setup menus.
Action:
Described per function. See Total, Set Points, Peak Hold, Tare
Memory.
SELECT
Function:
Used to enter numeric values and select stored Tare values and to
scroll through the setup menus.
Action:
Described per function. See Setup, and Total.
ENTER
Function:
Used to enter numeric values, setup parameters, and modify stored
Tare values.
Action:
Described per function. See Setup, Set Points, Tare Memory, and
Total.
SYSTEM INITIALIZE
This function should only be used when the operator is prepared for a
complete setup and Calibration procedure. Under no circumstances
should this procedure be initiated by the end user.
Function:
Clears the internal calibration settings. Starts the scale from the
beginning. Usually used for board replacement, troubleshooting, or
load cell replacement. This procedure does not alter factory features
settings.
Action:
To initiate a system initialization, push the SHIFT key and then push
the CALIBRATION key (inside the seal port). The message will
display “RESET”. A full calibration must follow this operation.