CALIBRATION PROCEDURE

1) Select

Channel Calibration

press 1.















2)

Start the Calibration procedure by pressing 1.


.




3) Press

1 to perform a completely new calibration.
















4)

The MSI-9000 CellScale prompts for the calibration units of mass. Press number value
corresponding to desired units.









***** Channel Setup *****

(1)

Channel Calibration

(2)

Data backup

(3)

Filter (software)

(4)

Channel Scanlist functions

(5)

Motion Compensation (Accelerometer option required)

(6)

Motion Detection

(7)

Tare

(8)

Total

(9)

Units Lock

(a)

Zero

(b)

Standard Mode: Industry

(c)

4 to 20mA

(d)

Temperature Sensor Installed: No

(ESC)

Exit

***** Choose Channel Parameters *****

(1)

Continue. These settings are correct

(2)

Choose the channel to calibrate:

1-1

(3)

Calibration

timeout:

None

(ESC)

Exit

***** Channel Calibration *****

(1) Perform a completely new calibration

(2) Re-do the current calibration
(3) Manually adjust current cal points
(4) Reset all cal parameters to defaults
(5) Change display resolution countby
(ESC) Exit

Calibrate channel #:

1-1

The channel name is:

....................

Calibration Capacity:

0.

Calibration Countby:

1.

Calibration Units:
Display Resolution Countby:

1.

Display Units:

***** Choose Calibration Units *****

(1)

Pounds

(

2)

Kilograms

(3)

Tons

(4)

Metric Tons

(5)

Ounces

(6)

Grams

(7)

Deka-Newtons


Type a key...