Torbal AGS Moisture Analyzer Instruction Manual User Manual

Page 44

Advertising
background image

44


Cleaning the Analyzer

• Before cleaning the unit or replacing halogen bulbs, always unplug the power cord from the electrical

outlet.

• Before performing any maintenance on the unit, make sure all parts and surfaces of the Analyzer have

cooled.

• Use a soft, lint-free, slightly damp cloth to clean the Analyzer.

• Wipe the unit gently. Do not allow any liquid to enter into the Analyzer’s weighing or drying

mechanisms.

• Do not touch the temperature sensor (see parts description).

• Do not apply extensive pressure to the LCD display.

• Do not use chemicals especially benzene when cleaning the surface. Corrosive chemicals may damage

the finish.

• Alcohol may be used only to clean the scale’s stainless steel pan support or the pan handle.

• To prolong the life of the halogen bulbs, be sure the bulbs are clean and free from fingerprints. Even a

small amount of oil from your fingers can cause the bulb to break when it is turned on.

Replacing the Halogen Bulb

Replace the halogen bulbs when they are defective or when you sense that drying is taking more
time than usual.

Warning:

• Before replacing the halogen bulb, avoid electrical shocks by unplugging the power cord from the

electrical outlet.

• Make sure the rated voltage of the new halogen bulbs is correct for the voltage of the power supply,
• Handle the old bulb with care to avoid cracking. Broken glass may cause injury.
• Never touch a halogen bulb with your bare fingers.
• Always wear gloves when handling a new bulb. Even a small amount of oil from your fingers can

cause the bulb to break (or shorten its life) when it is turned on.

• If a bulb is touched, before being used it may be cleaned with a lint-free cloth.
• When replacing the bulbs in the Analyzer, make sure to avoid touching the temperature sensor.

Chapter 20: Halogen Bulb Replacement and Analyzer Maintenance

Advertising