3

13

Connections and Assembly

Connecting Video Colposcope with Monitor

WARNING If peripheral devices (example: monitor, video printer, VCR ,other) do
not comply with IEC 60601-1-1, they must be kept out of the patient area (6 feet
minimum from patient).

WARNING All signal input and output (I/O) connectors are intended for
connection to only peripheral devices (example: monitor, video printer, VCR) that
are in compliance with IEC 60601-1 (General Requirements for Safety, Medical
Electrical Equipment) or other IEC standards (for example IEC 60950, Information
Technology Equipment - Safety) as appropriate to the nature of the peripheral
device. Connecting additional peripherals to the Video Colposcope may increase
the risk associated with chassis or patient leakage currents. To maintain operator
and patient safety, the User should consider the system leakage current
requirements of IEC 60601-1-1 (Medical Electrical Equipment, Safety
Requirements for Medical Electrical Systems). The user should measure leakage
currents accordingly to confirm that no electric shock hazard exists. An isolation
transformer that is in compliance with IEC 60601-1 used to power the additional
peripherals may be used to control the system leakage current to comply with the
requirements of IEC 60601-1-1.

LINE A

PARALLEL REMOTE RGB/COMPONENT

LINE B

~AC IN

IN

IN

IN

IN

IN

IN

IN

IN

IN

OUT

OUT

OUT

OUT

OUT

OUT

OUT

OUT

OUT

VIDEO

S-VIDEO

G/Y

B/Pb

R/Pr

VIDEO

AUDIO

AUDIO

AUDIO

S-Video Cable to Monitor
Line A S-Video (Y/C) In

Back Panel of Monitor