Rechargeable handle, Cleaning, Intermediate level disinfection – Welch Allyn Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, Accessories - Cleaning, Disinfection, and Sterilization - User Manual User Manual
Page 11
Advertising

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories
ENGLISH
9
Rechargeable Handle
Cleaning
1.
Using CaviWipes, wipe the surface to remove all gross debris.
2.
Using a soft bristle brush, scrub the handle for a minimum of 5 minutes then repeat
wipe with a second CaviWipe.
3.
Allow the handle surface to dry for a minimum of 10 minutes.
Intermediate Level Disinfection
1.
Follow the Cleaning instructions above.
2.
Using CaviWipes, wipe the handle and keep wet with CaviWipe solution for a
minimum of 5 minutes.
3.
Allow the handle surface to dry for a minimum of 10 minutes.
Rheostat
(Brightness
Control)
Part Number 71000-A
Caution
Do not saturate the rheostat area. Excess solution entering the
assembly may damageinternal components in the handle.