9

3. SUCTION MECHANISM

a) Connect suction tubing from an external

suction source to the suction nipple located
on the control body. Place the distal tip of the
endoscope in a basin of water and depress
the suction control valve. Water should be
rapidly aspirated into the suction system
collection container.

b) Release the suction control valve to deter-

mine if the valve freely returns to its OFF
position and the aspiration of water ceases.

NOTE: A rubber inlet seal in good condition must be
on the accessory channel inlet to prevent the loss of
suction. Worn seals will result in leakage and should
be replaced.

Suction
Tube

Depress

Suction Nipple

To Suction Source

Rubber Inlet Seal

WANM-3488-COL 4/19/99 9:27 AM Page 9