Cardioperfect workstation – Welch Allyn Cardioperfect Workstation, PEDMEANS ECG Interpretation Module - Physicians - User Manual User Manual
Page 19

CardioPerfect Workstation
PEDMEANS ECG Interpretation Module Physicians Manual
DIR 80015051 Rev. C
19 / 55
or
positive QRS amplitude in V3R, V1 > ULN
or
pure R or qR pattern in V1
and
positive QRS amplitude in 1 of V3R, V1 > ULN
or
negative QRS amplitude in V6 > ULN
or
RS ratio in V6 < 1.0
or
right axis deviation
or
moderate strain in 1 of V3R, V1, V2
or
pure R or qR pattern in V3R
and
RS ratio in V1 > 1.0
and
positive QRS amplitude in 1 of V3R, V1 > ULN
or
negative QRS amplitude in V6 > ULN
or
RS ratio in V6 < 1.0
or
right axis deviation
or
moderate strain in 1 of V3R, V1, V2
Say:
“possible RVH”
if:
positive QRS amplitude in V1 > ULN
and
right axis deviation
or
negative QRS amplitude in V6 > ULN
and
RS ratio in V6 < 0.66
and
right axis deviation
or
upright T in V1,V6
and
3 days < age < 6 years
or
pure R or qR pattern in 1 of V3R, V1
or
positive QRS amplitude in V1 > ULN + 200 µV
or
negative QRS amplitude in V6 > ULN + 200 µV
or