Battery led indicator, Battery care, Battery charging – Welch Allyn Cardioperfect PRO ECG Recorder - User Manual User Manual

Page 19: Storing the battery, Cardioperfect, Pro ecg recorder - user manual, Warning

Advertising
background image

CardioPerfect

PRO ECG Recorder - User Manual

DIR 80012763 Ver.

F

19 / 26

NOTE


The CardioPerfect Pro cardiograph can be powered by a USB connection or through a battery
pack. If you are using a USB connection (dummy battery pack) you do not have to charge the
battery pack. If a battery pack is used with a RS232 connection, you have to charge the
battery pack with a battery charger supplied by Welch Allyn.

Battery LED Indicator

The battery pack is provided with a (green) LED indicator that indicates the performance
status of the battery pack, the following applies:

LED Intermittent

– battery pack is charging after complete discharge, it is recommended to

not use the battery pack for use with the cardiograph for diagnostic purpose.
LED On

– battery pack is charging and not completely charged but can be used with the

cardiograph for diagnostic purpose.
LED Off

– battery pack is completely charged and ready to be used with the cardiograph for

diagnostic purpose.

Battery Care

A fully charged battery will provide a minimum of 8 hours of operation.

The CardioPerfect Pro cardiograph can be powered by a battery pack. This battery pack
contains rechargeable batteries. Battery performance is maintained when recharging takes
place when the battery is almost empty. This can be seen on the recorder when the LED on
the front lights up in red. This is the best time for recharging the battery pack. After the light
has turned red you have still enough power for about ten minutes of recording.

Both the internal gauge measurement and the percentage of battery capacity shown in Welch
Allyn CardioPerfect Workstation become less accurate.

Battery Charging

1. Pull the battery compartment from the recorder by holding the recorder with one hand

and pressing on both sides off the battery compartment with your other hand.

2. Connect the battery to an electrical outlet using the power adapter by inserting the

charger cable into the top of the battery compartment and charge your battery.

3. Click the battery compartment back into place after the battery is recharged.



Warning



Charging must be done outside of the patient environment.


Storing the Battery

If the cardiograph will not be used for more than three months, remove the battery pack from
the unit and place it into storage. While in storage, charge the battery to retain the battery life
and capacity. This ensures that the battery does not completely discharge while in storage.

Advertising