Welch Allyn CP 150 spirometry option - User Manual User Manual

Page 44

Advertising
background image

Symptom

Possible cause

Suggested action

Values are too high (consistently)

Pressure connection is partially

obstructed

Remove any foreign substance from the

flow transducer or pressure tubing.

Predictive values are blank

The selected norm does not support

certain values, and composite norm

values are disabled

Re-enter age/birthdate, height, gender,

race.
(Fill in the fields. All mandatory fields

must be filled in before you can

proceed.)

Enable composite norm values. See

Viewing or changing the spirometry

settings.

The flow sensor has been dropped.

Accident

Recalibrate. See Performing a

calibration.

Report does not print parameters or

graphs.

Improper parameter settings

Check print settings. See Viewing or

changing the spirometry settings.

Patient test values differ from values

expected by physician.

Various

If the transducer is contaminated with

sputum or secretions, replace it.

Verify that proper barometric pressure

has been entered. See Performing a

calibration.

Verify the patient data.

Eliminate any leaks in the pressure

tubing.

Retest using a nose clip.

Replace the sensor if damaged.

Recalibrate.

Replace the transducer and retest.

40 Troubleshooting

CP 150 spirometry option