Caster base accessory assembly instructions, Detachment instructions, Table clamp accessory assembly – Welch Allyn LS-150 Halogen Exam Lights - User Manual User Manual

Page 2: Specifications approximate physical dimensions, Environment, Transport/storage: operating, Service information, Ordering information, Floor stand models only), Table clamp accessories (wall mount models only)

Advertising
background image

CASTER BASE ACCESSORY ASSEMBLY INSTRUCTIONS

NOTE: The Exam Light must be assembled before the Caster Base Accessory is attached.

Attachment Instructions:

Place styrofoam block from accessory packaging box on floor. Then place Caster Base Accessory on
center of block so that casters are not touching the floor.

Position the Exam Light flexible arm in neutral position as shown in Figure 1.

Place the Exam Light base into Caster Base Accessory centering power cord between two wire
frame legs, as shown in Figure 2. The Exam Light base should rest on top of bent legs on the
opposite side.

Place hand and foot as shown in Figure 3. Use your foot to push the Exam Light base down firmly
until the base engages into the wire frame.

NOTE: Do not force base into place by pushing down on the Light pole or pulling on flexible arm.

Check to insure that the Exam Light is assembled correctly with the Caster Base Accessory. To do this, lay the Exam light over on its
side and verify that the Caster Base Assembly’s wire frame is interlocked into the notches of the Exam Light Base (see Figure 4). If
they are not interlocked, grasp the Exam Light Base with one hand and one leg of the Caster Base Accessory with the other hand,
and turn in opposite directions until the Caster Base Accessory locks into the Exam Light Base notches.

Detachment Instructions:

Place locking casters in locked position.

Place foot on one of the five legs of the metal frame. Then hold pole at the top near flexible arm and pole connection.

Simultaneously, lift and push the LS-135 & LS-150 Exam Light away from you, releasing the base from the Caster Base Accessory.

NOTE: Do not place foot on base while pushing and pulling. Do not pull on flexible arm.

When ordering replacement parts for your Caster Base Accessory, please use the following numbers:

TABLE CLAMP ACCESSORY ASSEMBLY

1.

Securely tighten the clamp (A) to the chosen mounting surface.

2.

Align the holes in the mounting bracket (B) on the light with the holes
in the table clamp.

Using the wrench provided, install the four screws provided and tighten.

SPECIFICATIONS

Approximate Physical Dimensions

The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within
the 89/336/EEC Electromagnetic Compatibility Directive.

Authorized European Representative Address:
European Regulatory Manager
Welch Allyn, LTD.
Navan Business Park,
Dublin Road
Navan, County Meath, Republic of Ireland
Tel: +353 46 90 67700
Fax: +353 46 90 67755

SERVICE INFORMATION

For Technical Support or to obtain return instructions, please contact your nearest Welch Allyn service center listed below:

ORDERING INFORMATION

DISASSEMBLY INSTRUCTIONS

When not in use, wrap the power cord and use a twist tie to hold it together. Place the power cord on top of the base.

To return a floor stand unit for service:

1.

Turn the power switch to the OFF position (

) and unplug the cord from the outlet.

2.

Using a small flat head screwdriver, CAREFULLY pull back the tabs at the top of the pole slightly - just enough to
disengage them from the black plastic insert in the pole.

3.

Pull up on the bottom end of the flexible arm. Unplug the electrical connector and remove Luminaire-Flexible Arm
assembly from the pole.

4.

Tilt the pole/base onto its side and unscrew the large phillips head screw located on the bottom center of the base.

5.

Remove the metal base cover plate and disconnect the connector attached to it.

6.

Disconnect the remaining electrical connector from the base and pull the pole out of the base.

7.

Securely wrap the three components (base, pole, and Luminaire assembly) and loose parts in suitable protective
packaging for shipment to your authorized service center.

To return a wall mount unit for service:

1.

Turn the power switch to the OFF position (

) and unplug the cord from the outlet.

2.

Pull up on transformer to remove it from its mounting screws.

3.

Unscrew the four screws mounting the unit to the wall. For units fitted with the table clamp accessory, simply
remove clamp from the table.

4.

Securely wrap the components (lamp assembly and transformer) in suitable protective packaging for shipment to
your authorized service center.

NOTE: This product contains no materials classified as hazardous. Therefore, its disposal is not deemed as
harmful to the environment or health risk to individuals disposing of the product. As a precaution, it is
recommended you contact your local disposal and/or recycling authority for information regarding the disposal of
the equipment.

Figure 1.

Flexible arm in
neutral position

Figure 2. Base fitted into two bent legs

Figure 3. Push base into place with foot.
Do not force using pole.

Figure 4. Caster Base Accessory assembled
correctly with Exam Light Base.

Washers

Caster Base Frame

440094 Caster with brake

OR

440095 Caster without brake

No. 440094 Caster with brake, Acorn Nut and two Washers

No. 440095 Caster without brake, Acorn Nut and two Washers

USA

+ 1 315 685 4560
800 535 6663

Australia

+ 61 29 638 3000
800 074 793

Canada

800 561 8797

China

+ 86 216 327 9631

European Call Center

+ 353 46 906 7790

France

+ 331 6009 3366

Germany

+ 49 747 792 7186

Japan

+ 81 33 219 0071

Latin America

+ 1 305 669 9003

Netherlands

+ 31 15 750 5000

Singapore

+ 65 6419 8100

South Africa

+ 27 11 777 7555

United Kingdom

+ 44 207 365 6780

Sweden

+ 46 85 853 6551

Catalog #

Description

04450

Halogen replacement lamp for LS-150 Halogen Exam Light

04430

Halogen replacement lamp for LS-135 Halogen Exam Light

44350

Caster base accessory, LS-135 & LS-150 Halogen Exam Light
(Floor Stand Models Only)

44101

Table Clamp Accessories (Wall mount Models Only)

44301

Rail Mount Accessory (Wall mount Models Only)

Ø 36 cm
(Ø14”)

89 cm
(35”)

63 cm
(25”)

9.5 cm
(3.75”)

12°

Leakage Current: The products listed within this manual comply with
the agency requirements listed below.

Environment:

Transport/Storage:

Operating:

-20°C - 49°C,

10°C - 35°C

10% - 95% R.H. Max,

30% - 75% R.H. Max,

500hPa - 1060hPa Altitude

500hPa - 1060hPa

Altitude

Safety Class I Equipment
Continuous Operation

ETL listed:
UL2601-1,
CSA C22.2 No. 601.1,
IEC 60601-1, AS 3200.1.
IEC 60601-1-2

Australia EMC Framework Compliance

23 cm
(9”)

86 cm
(34”)

C

US

N344

Model

Input

Output

Fuse

(3)

Wattage

Intensity

Spot Size

@ 65 cm

(24”)

Working

Distance

Color

Temperature

Lamp

Life

Total

Weight

Cord

Length

44300

120v~60Hz,

400mA

12v~2.9A

250v, 500mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

7.8 kg

(17.3 lbs)

2.9 M

(9.5’)

44302

230v~50Hz,

250mA

12v~2.9A

250v, 315mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

7.8 kg

(17.3 lbs)

2.9 M

(9.5’)

44303

100v~50/60Hz,

500mA

12v~2.9A

250v, 630mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

7.8 kg

(17.3 lbs)

2.9 M

(9.5’)

44304

230/240v~50Hz,

250mA

12v~2.9A

250v, 315mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

7.8 kg

(17.3 lbs)

2.9 M

(9.5’)

44306

240v~50Hz,

250mA

12v~2.9A

250v, 315mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

7.8 kg

(17.3 lbs)

2.9 M

(9.5’)

44310

120v~60Hz,

400mA

12v~2.9A

250v, 500mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

3.3 kg (7.3lbs)

2.4M (8’)

44312

230v~50Hz,

250mA

12v~2.9A

250v, 315mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

3.3 kg (7.3lbs)

2.4M (8’)

44313

100v~50/60Hz,

250mA

12v~2.9A

250v, 630mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

3.3 kg (7.3lbs)

2.4M (8’)

44314

230/240v ~50Hz,

500mA

12v~2.9A

250v, 315mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

3.3 kg (7.3lbs)

2.4M (8’)

44316

240v~50Hz,

250mA

12v~2.9A

250v, 315mA

35w

525fc (1)

29 cm

(11.4”)

30.5 cm min
(12” min) (2)

3000K

5,000 hrs (4)

3.3 kg (7.3lbs)

2.4M (8’)

44500

120v~60Hz,

600mA

12v~4.2A

350v, 800mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

7.9 kg

(17.4 lbs)

2.9 M

(9.5’)

44502

230v~50Hz,

350mA

12v~4.2A

250v, 500mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

7.9 kg

(17.4 lbs)

2.9 M

(9.5’)

44503

100v~50/60Hz,

750mA

12v~4.2A

250v,

1000mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

7.9 kg

(17.4 lbs)

2.9 M

(9.5’)

44504

230/240v~50Hz,

350mA

12v~4.2A

250v, 500mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

7.9 kg

(17.4 lbs)

2.9 M

(9.5’)

44506

240v~50Hz,

350mA

12v~4.2A

250v, 500mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

7.9 kg

(17.4 lbs)

2.9 M

(9.5’)

44510

120v~60Hz,

600mA

12v~4.2A

250v, 800mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

3.4kg

(7.4 lbs)

2.4M (8’)

44512

230v~50Hz,

350mA

12v~4.2A

250v, 500mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

3.4kg

(7.4 lbs)

2.4M (8’)

44513

1000v~50/

60Hz,750mA

12v~4.2A

250v,

1000mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

3.4kg

(7.4 lbs)

2.4M (8’)

44514

230/240v ~50Hz,

350mA

12v~4.2A

250v, 500mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

3.4kg

(7.4 lbs)

2.4M (8’)

44516

240v~50Hz,

350mA

12v~4.2A

250v, 500mA

50w

576fc

33 cm

(13”)

30.5 cm min
(12” min) (2)

4700K

3,000 hrs (4)

3.4kg

(7.4 lbs)

2.4M (8’)

(1)

Full intensity

(2)

Recommended minimum working distance, see cautions regarding skin exposure.

(3)

Non-user serviceable in-line single acting type.

(4)

Lamp life may vary slightly due to usage patterns.

80012404 Ver. A

(English)

Advertising
This manual is related to the following products: