Using the compacvideo otoscope system, Lamp replacement – Welch Allyn 23120P (PAL) CompacVideo Otoscope - User Manual User Manual

Page 10

Advertising
background image

5

Using The CompacVideo Otoscope
System

The CompacVideo Otoscope System allows the visual anatomy to be
shared with a patient. Indications for use include:

Inspection of the outer ear canal

Inspection of the tympanic membrane

Inspection of hearing aid devices

The CompacVideo Otoscope System should not be used for any other
purposes. For a Video System capable of operating multiple systems
(such as otoscopes, sinuscopes, rhinolaryngoscopes, and other
endoscopes), please contact Welch Allyn Customer Service and ask for
information on the Welch Allyn VDX-300.

Lamp Replacement

(Caution: The lamp base can get very hot. Be certain the lamp has
cooled before attempting to remove the lamp assembly.)

The CompacVideo Otoscope System uses Welch Allyn lamp # 03500.
Average lamp life is 20 hours.

To replace the lamp, remove the Video Otoscope Head from the power
handle. From the bottom of the Video Otoscope Head, grasp the base of
the lamp and pull straight out. A forceps or needle nose pliers may be
needed to firmly grasp the black lamp base.

Insert the new lamp assembly by pushing the new lamp assembly
firmly into the Video Otoscope Head. Attach the power handle to
firmly seat the lamp. Avoid removing the lamp for reasons other than
replacement.

Advertising
This manual is related to the following products: