AUDIBLE SIGNALS AND LIGHTS FOR OPERATION AND ALARM:

MAIN . . . . . . . . . . . . . . . . . . . . . . . . . . . .green light

ALARM, PATIENT PLATE CONTINUITY . . .low pitch intermittent audible alarm

- red light

ALARM, OUTPUT POWER . . . . . . . . . . . .higher pitch intermittent audible alarm

CUT AND BLEND MODES . . . . . . . . . . . .low pitch audible signal - yellow light

COAGULATION MODE . . . . . . . . . . . . . . .high pitch audible signal - blue light

NOTE: Specifications subject to change.

OUTPUT POWER AT 500 OHMS

PURE CUT . . . . . . . . . .100 W RMS (open circuit 830 Vp-p; crest factor 1.43)

Waveform: sinusoidal at 450 KHz

BLEND CUT . . . . . . . .100 W RMS (open circuit 1200 Vp-p; crest factor 2) Wave-

form: sinusoidal at 450 KHz
Duty cycle: 60%

COAGULATION . . . . . .80 W RMS (open circuit 3800 Vp-p; crest factor 5.5)

Waveform: sinusoidal at 450 KHz
Duty cycle: 10%

Output Power Diagrams according to PAR.6.8.3 (IEC 601-2-2)

Tolerance: 20% according to PAR.50.2 (IEC 601-2-2)

II.

SYSTEM FEATURES (cont.)

12

Electrosurgical Unit

• Instruction Manual