Welch Allyn 4607x Solid State Portable Headlight System - User Manual User Manual

Components, Set-up and operation, Maintenance

Advertising
background image

Symbol Descriptions

Warnings & Cautions

Guidelines for the safe use of the Solid State
Portable Headlight

Introduction

Thank you for purchasing the Solid State Portable Headlight by Welch Allyn,
the worldwide leader in illuminated medical diagnostic instrumentation. By
following the use and care guidelines supplied in this owner’s manual, you
will enjoy years of dependable, trouble-free service from your new
headlight.
This headlight is designed to meet the various needs of a physicians’ office,
hospital environment, and wherever compact, portable, direct illumination
is critical.

Components

Standard Parts List

Solid State Portable Headlight and Power Cord/Charger

Accessory Parts List

Replacement Parts List

When ordering replacement parts for your Solid State Portable Headlight,
please use the following numbers

Set-up and Operation

Instructions for Use

The Solid State Portable Headlight (REF 46070), consisting of Portable
Headlight with soft terry cloth headband (REF 46000) and charger/direct
power source (REF 74170). The Solid State Portable Headlight is designed
to operate off the battery or direct power source. The headlight is provided
fully assembled to operate in the portable operation mode. All headlights
are shipped with the batteries discharged
. Charge the batteries prior to
use. Once the unit is charged, turn the power switch, located at the top of
the luminaire, to the on position. To operate the headlight using the direct
power source, insert the charger/direct power source (REF 74170) into the
input port located on the top of the luminaire (see photo below). To charge
the battery pack, connect the charger/direct power source (REF 74170) into
the top of the luminaire (see photo below). To adjust the spot size, rotate
the front section of the luminaire. To adjust the headband, simply move the
Velcro adjustable headband to a comfortable position.
Use the Solid State Portable Headlight to provide illumination while
performing a number of activities in the physician’s office, hospital, or
emergency environment such as minor suturing, general exam, and foreign
body removal.

Maintenance

Each month examine cords and connections. The user can perform battery
and/or headband replacement on an as needed basis. Return the unit to an
authorized Welch Allyn Technical Service Department or distributor for
additional servicing. See “Service and Warranty Information” section.

Battery Charging

The fully drained battery will reach a 90% charge retention in approximately
three hours of continuous charging. This is approximately 25% longer if the
charging is happening at a temperature less than 15° C. Proportional usage
will require a proportional charge time such as if the product is used for 15
minutes, charge the battery for 45 minutes to reach a 90% capacity.
Overnight charging brings the battery to its 100% full capacity.
Charging the unit while operating will significantly increase the charge time
to over 12 hours to obtain a 90% charge level.
This battery pack does not exhibit any memory problems that exist with
NiCad batteries. However, like other products where batteries are used,
remove the battery from the product when the product is in storage or
when not in use for extended periods beyond 6 months. If the battery is left
in the product for extended periods of time, the battery will drain down to
an unusable level and require charging prior to use.

Battery Replacement

Headband Replacement

Cleaning

(Recommended monthly)

1.

Unplug the unit prior to cleaning.

2.

Wipe down the entire luminaire with a cloth slightly dampened
with a mild solution of detergent. The terry cloth headband is
removable and is machine washable.

3.

Do not operate the light until the light is thoroughly dry.

CAUTION: Do not immerse in cleaning solutions.

Troubleshooting

Specifications

Electrical

Attention: Consult user’s Manual for additional
information

Double Insulation

INTENSE LIGHT

:

Lamp is bright. Do not stare at the

lamp when in use.

May be harmful to eyes.

Lamp On

ATTENTION: Read Operating Manual for Warnings,
Cautions, and Instructions for Use.

IPXØ: Not protected against the ingress of water.

WARNING: Solid State Portable Headlight is not suitable
for use in the presence of a flammable anesthetic
mixture with air or oxygen or nitrous oxide. An explosion
may result.

INTENSE LIGHT

:

Lamp is bright. Do not stare at the

lamp when in use.

May be harmful to eyes.

USE ONLY:
Solid State Portable Headlight Power Source/Battery
Charger and rechargeable battery pack supplied by Welch
Allyn.

There are no user-serviceable parts inside the device
except the rechargeable batteries and headband. Contact
the Authorized Service Center listed at the end of this
document for further assistance.

Solid State Portable Headlight System

REF 46070, 46072, 46074, 46076

Part No. 460132 Rev. B

Lamp Off

General practices to minimize risk of harm to the skin
from optical radiation hazards include: minimizing
illumination intensity at the tissue examination site,
minimizing exposure times, and taking additional
precautions when skin sensitivity has been altered
through tissue trauma or the use of anesthesia.
General good practices to minimize risk of harm to the
eyes from optical radiation hazards include: Avoiding
looking at bright light sources and their reflections, and
protect eyes where normal pupil sizes and aversion
responses are not present.

Model 46070 (Domestic)
Model 46072 (European)
Model 46074 (U.K.)
Model 46076 (Australia)

Model 05232-U Soft Case

Model 49642 Soft Terrycloth
Headband – case of 10

Model 72270 Rechargeable
Battery

Direct Power Source/Charger:
Model 74170 Domestic
Model 74172 European
Model 74174 U.K.
Model 74176 Australia

Follow the assembly instructions listed below and the
corresponding photographs.

.

Rotate to adjust
spot size.

On/Off Switch

Input for Charger/
Direct Power Source

Headband Adjustment

Drop the battery pack
into the luminaire visor
and plug into the
connector.

Excess solution entering the luminaire assembly could damage
internal components. Use caution to ensure cloth is not
saturated with solution.

Do not sterilize the unit.

Trouble

Area

Possible Cause

Corrective Action

No Light
Output

Dead battery or
battery not connected
properly

Verify that the unit operates on
direct power source. If yes, insure
that the battery is connected
properly; refer to the battery
replacement section for instructions
on connecting the battery. If the
battery is connected properly,
recharge or replace battery

Low
Light
Output

Low charge on battery

Verify that the unit operates on
direct power source. If yes,
recharge or replace battery

Ref

Model

Input Voltage

Output

Voltage

Leakage Current

46070

120VAC ± 50/60 Hz

5 volts vdc

<10 uA

46072

230VAC ± 50/60 Hz

5 volts vdc

<10 uA

46074

230VAC ± 50/60 Hz

5 volts vdc

<10 uA

46076

240VAC ± 50/60 Hz

5 volts vdc

<10 uA

On/Off Switch

Holding the luminaire
with the on/off switch in
the up position, place
the visor section of the
luminaire into the
openings located in the
front section of the terry
cloth headband.