Symbol descriptions, Indications, Contraindications – Welch Allyn Stryker Light Source For Q2002-00, Q2002-02, Q2002-04, Q2002-06 - User Manual User Manual
Page 5

1-1
Symbol Descriptions
Attention: Read Operating Manual for Warnings, Precautions,
and Instructions for Use.
Dangerous Voltage: Refer Servicing to Qualified Personnel.
Type B Equipment
Caution — Hot
Warning: Risk of Fire. Replace Fuses as Marked.
Light; Lighting; Illumination Intensity
Indications
Use of the Stryker Light Source is indicated for medical lighting
applications requiring a remotely located, high-quality white light. Only
qualified mecial personnel may use this product.
Contraindications
The use of the Stryker Light Source is contraindicated for applications
where high intensity light might burn or damage tissue; such as for
neonate transillumination and ophthalmic procedures. For further
contraindications, the user must also consult the operating manual of the
instrument in use with the Light Source in the medical procedure.