Trouble shooting guide – Welch Allyn 73210 6V Power Supply Rectal Light Handle - User Manual User Manual
Page 16

12
Trouble Shooting Guide
Problem
Possible Cause
Solution
Lamp does not
light.
Rectal Light Handle is not
plugged into the power
supply.
Verify that the Rectal Light Handle is
properly plugged into power supply
cord.
Power Supply is not plugged
in.
Verify the power supply is properly
plugged into the outlet.
Faulty Lamp
Replace with a new lamp.
Loose Lamp
Verify that the lamp is secured into the
Rectal Light Handle.
Wrong Lamp
Verify that the lamp is a REF 07800
Welch Allyn lamp.
Damaged Lamp
Inspect the lamp and the o-ring for
damage. Replace with REF 07800
lamp if damage is found.
Rectal Light
Handle does not
stay plugged into
the transformer
cord.
Rectal Light Handle/
transformer interconnect
plug is damaged.
Inspect interconnect plug for damage.
Replace either or both REF 73210 or
REF 73305, 73322, 73324, 73326.
Light output
seems low.
Faulty Lamp
Replace with a new lamp.
Dirty lamp lens
Inspect lamp lens for debris and clean.
This guide is intended to provide assistance towards determining the fault condition of
this product. Please contact the authorized Welch Allyn Service Center listed within the
product owners manual for additional assistance.