Welch Allyn Allegiance Brand Tactics Blood Pressure Cuffs - User Manual User Manual

Allegiance brand tactics blood pressure cuffs

Advertising
background image

For 30502-series and 30505-series
Blood Pressure Cuffs and 30502-series and
30505-series Aneroid Sphygmomanometers

Setup
Applying Cuff
1. Seat the patient and make them as relaxed and comfortable as pos-
sible. For arm readings, ensure the patient’s upper arm is at heart level
and that the lower arm is passively supported.
2. Select the cuff size appropriate for limb.
NOTE: Verify that the index line on the cuff falls within the range area
indicated on the cuff. If the index line falls short or is past the range, use a
larger or smaller cuff, respectively, to ensure accurate results. If the mark-
ings on the cuff are not visible, do not use the cuff.
3. Wrap the cuff around the bare limb with the “artery” arrow located
over the brachial artery with the lower border about 2.5cm above the
antecubital space of the arm.

Operation
Auscultatory Method – Manual Device (not appropriate for
neonatal patients)
1. Place the earpieces of the stethoscope firmly in the ear
canal. Place the bell side of the chest piece over the brachial
artery at the antecubital space.
2. Rapidly inflate the cuff to a level that is about 30 mm Hg
above the estimated (or palpatory) systolic pressure.
3. Partially open the valve to allow deflation at the rate of 2 to
3 mm Hg per second. Listen for the systolic pressure at the
first appearance of repetitive sounds during deflation. (Phase 1).

1

4. Listen for the diastolic pressure at the point when the Korotkoff
sounds disappear, as the cuff deflates (Phase V).

2

5. Rapidly release the remaining pressure and record the measurements
immediately. After a minimum of 30 seconds, repeat the steps above
for a second reading.
Oscillometric Method (Electronic Device)
1. Check device instructions for compatibility with cuff.
2. Insure the monitor is in the correct mode (adult, pediatric,
neonatal etc.).
3. Connect the cuff connector(s) to the device adapter(s).
4. Follow device instructions for taking blood pressure.

Two-tube cuffs with clamp
To make a two-tube cuff with clamp into a one-tube cuff, depress the
clamp to block off airflow through the tube.

Cleaning
Gauge: Clean the manometer with a slightly dampened cloth or
alcohol wipe.
Blood Pressure Cuffs: Clean the cuffs with a cloth dampened with soap or
detergent. Allow the cuff to air dry.
Disinfection and Sterilization: Do not use steam, heat or liquid disinfec-
tants to sterilize the cuff, bulb or tubing. Use gas sterilization if necessary.

DO NOT PRESS WITH HOT IRON

D. Perloff et al., Human Blood Pressure Determination by Sphygmomanometry
Dallas: 1994, 16-17.
Ibid., 17.

6/07

95P476 Rev. D

Allegiance Brand Tactics Blood Pressure Cuffs

TM

R

3.5”W x 8.5”H Insert. Do not alter art or size.
All prints in 100% Black.

3.5”

1

2

Advertising
Table of contents