Directions for Use
Chapter 2 Assembly and Operation
15
Concentrated high intensity light energy may result in thermal
injury to the tissue.
The risk of injury is increased under the following conditions:
•
Prolonged close stationary illumination of tissue.
•
Using a high intensity light source at levels greater than
needed for good visualization.
•
Using multiple instruments to illuminate the same area.
The following recommendations will reduce the risk of thermal
injury:
•
Use the minimum level of illumination necessary for
adequate visualization.
•
When possible, avoid close stationary viewing.
•
Minimize exposure time.
WARNING Do not use headlight at distances closer
than 10 inches (25 cm). Tissue damage or erythema may
result.