Welch Allyn Smartlink Event Pro SW Version: 4.10 Is Releasing and Shipping Addendum - User Manual User Manual

Addendum to the directions for use, Specifications: electromagnetic compatibility

Advertising
background image

Addendum to the directions for use

Propaq CS Vital Signs Monitor 810-1441-XX, 810-2734-00

Addendum to the directions for use

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

Addenda à la documentation d’utilisation

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

Supplemento a istruzioni per l'uso

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

Addendum zum Gebrauchsanweisung

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Anexo de instrucciones de uso

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Adenda às instruções de utilização

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

Addendum bij de Gebruiksaanwijzing

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

Tillägg till bruksanvisning

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

Käyttöppaan liite

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

Uzupełnienie instrukcji użytkownika

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

Addendum to the directions for use

For information about cleaning the blood-pressure cuff, see the cuff manufacturer’s instructions.

All Sure CO

2

accessories have been replaced with Surestream CO

2

accessories. (See Welch Allyn

publication 810-0409-XX, Products and Accessories.) Surestream is a trademark of Oridion
Systems, Ltd.

Addendum to the directions for use (SW version 3.7)

This section supplements 810-2734-00A, the multilanguage CD of Propaq CS monitor user manuals.

Specifications: Electromagnetic compatibility

Guidance and manufacturer’s declaration—electromagnetic immunity

The Welch Allyn Propaq CS monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of the Welch Allyn Propaq CS monitor should assure that
it is used in such an environment.

.

Immunity test

Test level

Compliance level

Electromagnetic environment - guidance

EN ISO9919

20 V/m. 1 KHz AM modulation

20 V/m

Intended for use during patient transport outside the healthcare
facility

EN ISO21647

20 V/m. 1 KHz AM modulation

3 V/m

Intended for use during patient transport outside the healthcare
facility