Set the date and time

After recharging a dead battery or after disconnecting the battery for a few minutes, you
must program the date and time screen.

You can change several device operating parameters in the Internal Configuration Mode.
When changed, these settings become the default power-up settings. You will also see
non-changeable device configurations for technical service purposes.

To enter the Internal Configuration Mode:

1.

Turn the Spot Vital Signs off (

).

2.

Press and hold the Blood Pressure Start/Stop (

) and Power (

) buttons. The

device enters the Internal Configuration Mode and displays the software version.

If the device displays E38, press the Next Patient/Clear/Cancel (

) button to

cancel the error.

3.

Press the Mode (

) button to cycle through the Internal Configuration menu until

you see the date option displayed on the screen.

4.

Use the Next Patient/Clear/Cancel (

) or Blood Pressure Start/Stop (

)

buttons to change the default setting.