Service and Maintenance Toolset

Connex ® Vital Signs Monitor 6000 Series™

The following is the listing of tools to perform service and maintenance, as well as basic functional verification checks,
full functional verification checks and calibration on the Welch Allyn Connex Vital Signs Monitor. Please refer to the
device service manual for detailed descriptions of what these activities provide. Note: These tools must also be readily
available to complete Welch Allyn Technical Service Training for this device.

Required Service Tools and Test Equipment*

Material

Number

Description

Quantity

Required

Usage

407691

VSM6000 SERVICE TEST BOX

1

407022

ASSY, MMF NURSE CALL CABLE – SERVICE

++

1

N/A

Welch Allyn Service Tool Software

1

4500-925

Connectivity Accessory Kit - USB Cable

1

6000-30

1-Tube Blood pressure hose (5 ft)

1

660-0138-00

CABLE, PATCH 5' RJ45,T568B

1

Required to test NIBP

407672

BP TEST VOLUME REPAIR FIXTURE

1

4500-30

Blood Pressure Hose (5')

1

620216

Fitting “Y” 1/8 x 1/8 x 1/8

1

200-2000IN

PRESSURE METER, NETECH

1

Required to test Thermometry (SureTemp or Braun)

01802-110

TESTER, CALIBRATION, 9600 PLUS

3

Required for Devices Equipped SureTemp Thermometry

06138-000

CAL-KEY, ASSY, M690/692

1

Required for Devices Equipped With CO2 Monitoring

T46530RF-2BD

CO2 Calibration Kit

1

Required for Devices Equipped With Masimo Acoustic Respiration Rate (RRa)

105460

Masimo RRa Tester

1

Required for Devices Equipped With Masimo SpO2

103507

TESTER, SPO2, MX RAINBOW

1

Required For Devices Equipped with Nellcor SpO2

DOC-10

CABLE, SP02 EXTENSION, NELCOR

1

SRC-MAX

NELLCOR PORTABLE OXIMETRY TESTER (MP506)

1


Indicates items only available from
Welch Allyn, or a Welch Allyn
distributor.

Indicates equipment required for full
functional verification checks and
calibration. The Welch Allyn Service Tool
software and an active Gold License is also
required to perform these checks. The
Gold License is obtained upon successful
completion of Welch Allyn Technical
Service Training.

Indicates equipment used for basic functional
verification checks. The Welch Allyn Service
Tool software (Silver Level) can be used to
perform the NIBP checks and SWF
upgrades/updates as well as installing
Licensable features on your device.**