Welch Allyn Propaq LT Monitor Radio Configuration - Quick Reference Guide User Manual

Radio configuration guide, Overview, Propaq lt

Advertising
background image

Radio configuration guide

Propaq LT

®

vital signs monitor

Overview

This guide explains how to do the following:

Configure your monitor radio card to your network.

Customize your radio configuration.

Reset your monitor radio.

Set your SSID.

It applies to new monitors or monitors returned from the Welch Allyn service department.
Your network must be configured to use factory-supported SSIDs and authentication and
encryption schemes.

This procedure is for Propaq LT “A” monitors (802LTAN and 802LTAS) with software
versions 1.60.XX or later.

To make any changes to your monitor radio card, access the monitor service menu. The
service menu is only available when the monitor is in adult patient mode.

To access the monitor service menu

From any main display screen (such as Large Numerics, Dual Waveform, ...):

1.

Highlight

, HR/PR, SpO2, NIBP, Resp, or

, and then press

.

2.

Highlight Setup (at the bottom of the screen) and press

. The setup menu

appears.

3.

Scroll right (

) to highlight Service, and press

. The service menu appears, with

NIBP and Show Info highlighted.

When in the monitor service menu, you can set the authentication and encryption
scheme and the network name through the monitor user interface.

Caution The instructions in this guide must be performed only by a qualified
biomedical service engineer or technician who is familiar with your facility’s
wireless network. If you are not familiar with your facility’s wireless network,
perform these tasks only with the assistance of a qualified technician.

Advertising