Welch Allyn PCH100 and PCH200 Starting a Test and Retrieving Data - Quick Reference Guide User Manual
Pch100, And pch200, Quick reference
006-700-0100-A
Reorder No. 706515
PCH100
and PCH200
Quick Reference
Caution — Consult accompanying documentation. Please see CD for translated versions of this card.
Starting a Test: Previewing the ECG (Wireless)
1.
Prepare the patient according to the HR100/300/1200 Directions for Use.
2.
Remove the battery door.
3.
Insert an SD card into the recorder.
4.
Insert a new battery (HR100) or batteries (HR300/1200).
5.
Start the Holter software application and select Test Startup.
6.
If the patient has a previous recording, click Search. A patient list displays.
•
Use the fields at the top of the patient list to filter your search.
•
Select a patient.
7.
Type or select all appropriate information.
8.
Click Next. The Recorder Selection
window will display.
9.
Wait for approximately 20 seconds for the serial number of the recorder to appear in the Recorders in
Range window.
10. Click Next. (If no number displays, click Troubleshooting.)
Lead
AHA Color
IEC Color
Ch1+
Ch1-
Ch2+
Ch2-
Ch3+
Ch3-
RL
Red
White
Brown
Black
Orange
Blue
Green
Green
Red
White
Yellow
Orange
Blue
Black