Welch Allyn CP200, Spirometry Option - Quick Reference Guide User Manual

Cp 200, Spirometry option quick reference

Advertising
background image

CP 200

™

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5

66

63

USA

· F

or

worldwide T

echnical S

upport numbe

rs

, s

ee

spirometry manual.

©2

005 · Printed in

USA · www

.welchal

lyn.com

· Mate

rial #

703977 Rev

B

CP 200

™

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5

66

63

USA

· F

or

worldwide T

echnical S

upport numbe

rs

, s

ee

spirometry manual.

©2

005 · Printed in

USA · www

.welchal

lyn.com

· Mate

rial #

703977 Rev

B

CP 200

™

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5 66

63

USA

· For

worldwide

T

echnica

l Sup

port numbers,

see spirometry

ma

nual.

©20

05 ·

Prin

ted

in

USA · www

.welchallyn.com · Material #

703

977

Re

v

B

CP 200

™

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5 66

63

USA

· For

worldwide

T

echnica

l Sup

port numbers,

see spirometry

ma

nual.

©20

05 ·

Prin

ted

in

USA · www

.welchallyn.com · Material #

703

977

Re

v

B

Advertising