Welch Allyn CP200, Spirometry Option - Quick Reference Guide User Manual

Cp 200, Spirometry option quick reference

Advertising
background image

CP 200

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5

66

63

USA

· F

or

worldwide T

echnical S

upport numbe

rs

, s

ee

spirometry manual.

©2

005 · Printed in

USA · www

.welchal

lyn.com

· Mate

rial #

703977 Rev

B

CP 200

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5

66

63

USA

· F

or

worldwide T

echnical S

upport numbe

rs

, s

ee

spirometry manual.

©2

005 · Printed in

USA · www

.welchal

lyn.com

· Mate

rial #

703977 Rev

B

CP 200

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5 66

63

USA

· For

worldwide

T

echnica

l Sup

port numbers,

see spirometry

ma

nual.

©20

05 ·

Prin

ted

in

USA · www

.welchallyn.com · Material #

703

977

Re

v

B

CP 200

Spirometry Option

Quick Reference

To Record an FVC Test

1. Connect spirometer to electrocardiograph.
2. Press .
3. Enter or search for patient data (age, height, gender, race)

.

4. Select effort and curve types.
5. Press Start. Perform effort. Press Stop.
6. Accept or reject effort. Repeat from Step 4 until finished

.

7. Press Test Done

.

To Calibrate the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings > Calibration Settings >

Calibrate Spirometer.

4. Enter transducer lot code and calibration code.
5. Update ambient temperature, humidity, and barometric pressure.

(See your local weather reports.)

6. Follow prompts

.

To Configure the Spirometer

1. Connect spirometer to electrocardiograph.
2. Press Menu key

.

3. Choose Spirometry Settings.
4. Change desired settings.

Caution — consult accompanying documentation.

80

0 53

5 66

63

USA

· For

worldwide

T

echnica

l Sup

port numbers,

see spirometry

ma

nual.

©20

05 ·

Prin

ted

in

USA · www

.welchallyn.com · Material #

703

977

Re

v

B

Advertising