Welch Allyn CP100/CP200 Battery Connection - Quick Reference Guide User Manual
Page 2
CP100 & CP200
2
6.
Carefully tuck in all cables. Place the battery into the battery compartment.
Fold the strap so that it will not stick out when you replace the battery door.
See
Figure 4
.
7.
Replace the battery door and screws. Return the electrocardiograph to its
normal, upright position.
The electrocardiograph turns on automatically and displays a prompt to enter
date and time.
8.
Enter the date and time.
9.
Verify the local configuration settings. See “Reviewing the Device
Configuration Settings” in the electrocardiograph DFU for further information.
10. Keep the electrocardiograph plugged in for at least 12 hours before using
it on battery power.
Figure 3. Plugging In the Battery Fuse
Figure 4. Folding the Battery Strap
“Z” wire
The electrocardiograph is ready to
use after the battery is fully charged.