Quick Reference Guide

Pre-test Set-up:

Push any button to turn the unit on.

Position the patient so that the test can be conducted level with, and square to, patient’s eyes.

Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double
beep and see the desired icon on the LCD.

Explain the test procedure to the patient:

“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”

Position yourself at eye-level and square with the face of the patient.

Test Procedure:

Push the “GO” button on the unit

Check that the patient is fixated correctly throughout the test.

Position the unit at the correct distance from the patient:

When the unit is too far away, you will hear slow, low-pitched beeps.

• • • • • •

Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———

When the unit is too close, you will hear quick, high-pitched beeps.

••••••

The crosshair will flash in synch with the tones.

Look through the peephole and align the crosshair on the pupil of the patient’s right eye. While the unit is acquiring data, you
will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °

If you are not acquiring data (steady tone without acquiring data chirps):

Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.

Ask the patient if they can see the red light.

When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not
need to press any buttons for the left eye test.)

Without changing position, rotate the unit to the left eye and align the crosshair over the pupil. Repeat test.

At the end of the test, you will hear the “tah-dah” sound again.

If the unit has not gathered enough good readings from either eye, you
will hear 5 tones when the test stops. - - - - - You must re-test that eye.

To stop a test at any time, hold any button until 5 tones sound.

Hints To Insure Good Results & Rapid Test:

Ambient lighting:

Don’t perform by uncovered windows

Dimmed light can help for those with small pupils

Too dim is difficult for those with dark irises

Entry angle - straight and level

Rotate unit to test left eye

Fixation: “Look at the red light”

Remind and monitor

Match patient age and mode

These are noted on the print out

Correct distance/Not taking readings (steady tone without chirps):

Ensure you are level and square with the patient

Scan the crosshair around the eye in an outward moving spiral until
chirps begin, hold this location

Ask patient if they can see the red light

Aim high if testing through glasses

Make sure the patient’s eyelids are not occluding the pupil

Quick Reference Guide

Pre-test Set-up:

Push any button to turn the unit on.

Position the patient so that the test can be conducted level with, and square to, patient’s eyes.

Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double
beep and see the desired icon on the LCD.

Explain the test procedure to the patient:

“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”

Position yourself at eye-level and square with the face of the patient.

Test Procedure:

Push the “GO” button on the unit

Check that the patient is fixated correctly throughout the test.

Position the unit at the correct distance from the patient:

When the unit is too far away, you will hear slow, low-pitched beeps.

• • • • • •

Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———

When the unit is too close, you will hear quick, high-pitched beeps.

••••••

The crosshair will flash in synch with the tones.

Look through the peephole and align the crosshair on the pupil of the patient’s right eye. While the unit is acquiring data, you
will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °

If you are not acquiring data (steady tone without acquiring data chirps):

Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.

Ask the patient if they can see the red light.

When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not
need to press any buttons for the left eye test.)

Without changing position, rotate the unit to the left eye and align the crosshair over the pupil. Repeat test.

At the end of the test, you will hear the “tah-dah” sound again.

If the unit has not gathered enough good readings from either eye, you
will hear 5 tones when the test stops. - - - - - You must re-test that eye.

To stop a test at any time, hold any button until 5 tones sound.

Hints To Insure Good Results & Rapid Test:

Ambient lighting:

Don’t perform by uncovered windows

Dimmed light can help for those with small pupils

Too dim is difficult for those with dark irises

Entry angle - straight and level

Rotate unit to test left eye

Fixation: “Look at the red light”

Remind and monitor

Match patient age and mode

These are noted on the print out

Correct distance/Not taking readings (steady tone without chirps):

Ensure you are level and square with the patient

Scan the crosshair around the eye in an outward moving spiral until
chirps begin, hold this location

Ask patient if they can see the red light

Aim high if testing through glasses

Make sure the patient’s eyelids are not occluding the pupil

TM

R

Left/Right Toggle & Print

• Selects eye to re-test

• Prints results (hold button down

while aiming patient side of
SureSight™ at printer until you
hear “tah-dah”)

Clear and Child/Adult
Calibration

• Clears readings

• Changes child/adult

(hold down)

“GO” Button

Unit on: starts test

Unit off: recalls old

readings

TM

R

Left/Right Toggle & Print

• Selects eye to re-test

• Prints results (hold button down

while aiming patient side of
SureSight™ at printer until you
hear “tah-dah”)

Clear and Child/Adult
Calibration

• Clears readings

• Changes child/adult

(hold down)

“GO” Button

Unit on: starts test

Unit off: recalls old

readings