Welch Allyn Auriflush Ear Wash System - Quick Reference Guide User Manual

Quick reference guide

Advertising
background image

Figure 2: Attaching an eartip to the handle (note that the tab
on the eartip is facing up).

Figure 3: Correct and incorrect irrigation techniques.
Arrows indicate water flow.

Ridge

Groove

Quick Reference Guide

Match
colors

Pull down
on ring

Snap Aerator

To Faucet

Injection
Port

Figure 1: Auriflush Ear Wash System Assembly

I n i t i a l S e t U p

See the manual for more detailed instructions.

1. Unscrew current aerator and its washer from the faucet.

2. Attach the snap aerator.

3. Match the colors and connect the tubing to the chamber.

4. Attach the chamber to the faucet. Pull down the white

plastic ring on the top of the chamber and then snap the
chamber over the snap aerator on the faucet. When the
chamber is in place, release the white plastic ring.

5. Set the temperature. First turn on the cold water

completely, then gradually turn on the hot water.

6. Depress the actuator on the handle. The blue sensor on

the handle will become white when the water tempera-
ture going through it becomes approximately 90° F. If the
water is too cold, the patient may experience vertigo or
discomfort.

NOTE: If the water temperature exceeds 110°F, water will stop

flowing from the eartip. Reduce the water temperature
and then adjust to an approximate body temperature
again.

NOTE: If you are experiencing an excessive amount of water

splash around your sink, you may attach the optional
chamber aerator to the bottom of the chamber. See
the Operating Instruction Manual for more details.

CORRECT FLOW

INCORRECT FLOW