Pre-test Set-up:

•

Push any button to turn the unit on.

•

Position the patient so that the test can be conducted level with, and square to, patient’s eyes.

•

Choose the appropriate mode using the child/adult button (child mode for 6 yr. and under). Hold button until you hear a double
beep and see the desired icon on the LCD.

•

Explain the test procedure to the patient:

•

“Now I will check your eyes. You will see a red light in the middle of the blinking green lights.”

•

Position yourself at eye-level and square with the face of the patient.

Test Procedure:

•

Push the “GO” button on the unit

•

Check that the patient is fixated correctly throughout the test.

•

Position the unit at the correct distance from the patient:

•

When the unit is too far away, you will hear slow, low-pitched beeps.

• • • • • •

•

Slowly move closer. At the correct distance (14”), you will hear a steady, low tone. ———

•

When the unit is too close, you will hear quick, high-pitched beeps.

••••••

•

The crosshair will flash in synch with the tones.

•

Look through the peephole and align the crosshair on the pupil of the patient’s right eye. While the unit is acquiring data, you
will hear a very high-pitched chirping sound over the steady low tone. ° ° ° ° °

•

If you are not acquiring data (steady tone without acquiring data chirps):

•

Scan around the pupil in an outward-moving spiral until chirps begin, then hold this location.

•

Ask the patient if they can see the red light.

•

When the test of the right eye is complete, you will hear the “tah-dah” sound. (Testing resumes after 1 second, so you do not
need to press any buttons for the left eye test.)

•

Without changing position, rotate the unit to the left eye and align the crosshair over the pupil. Repeat test.

•

At the end of the test, you will hear the “tah-dah” sound again.

•

If the unit has not gathered enough good readings from either eye, you
will hear 5 tones when the test stops. - - - - - You must re-test that eye.

•

To stop a test at any time, hold any button until 5 tones sound.

Hints To Insure Good Results & Rapid Test:

•

Ambient lighting:

•

Don’t perform by uncovered windows

•

Dimmed light can help for those with small pupils

•

Too dim is difficult for those with dark irises

•

Entry angle - straight and level

•

Rotate unit to test left eye

•

Fixation: “Look at the red light”

•

Remind and monitor

•

Match patient age and mode

•

These are noted on the print out

•

Correct distance/Not taking readings (steady tone without chirps):

•

Ensure you are level and square with the patient

•

Scan the crosshair around the eye in an outward moving spiral until
chirps begin, hold this location

•

Ask patient if they can see the red light

•

Aim high if testing through glasses

•

Make sure the patient’s eyelids are not occluding the pupil

TM

R

Left/Right Toggle & Print

• Selects eye to re-test

• Prints results (hold button down

while aiming patient side of
SureSight™ at printer until you
hear “tah-dah”)

Clear and Child/Adult
Calibration

• Clears readings

• Changes child/adult

(hold down)

“GO” Button

• Unit on: starts test

• Unit off: recalls old