Welch Allyn Connex Vital Signs Monitor Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 4

Species
Code Definition /
Description
Original / Alert
type
Traditional code
map
Details
Host
controller
action
6
6
Equipment alarm E11
Cuff pressure
duration
exceeded.
Reset Mod PG.
6
7
Equipment alarm Internal error
Internal NIBP
error. Contact
Welch Allyn
Technical
Service.
Power down.
6
8
Equipment alarm Internal error
Internal NIBP
error. Contact
Welch Allyn
Technical
Service.
Power down.
6
9
Host controller
alert
Internal error
None
Exit STAT
mode.
6
11
Host controller
alert
Internal error
None
None
6
12
Host controller
alert
Internal error
None
None
6
13
Host controller
alert
Internal error
None
None
6
22
Equipment alarm Internal error
Internal
malfunction.
Contact Welch
Allyn Technical
Service.
Power down.
This is only
available
during
software
loading.
6
23
Equipment alarm Internal error
Internal
malfunction.
Contact Welch
Allyn Technical
Service.
Power down.
This is only
available
during
software
loading.
6
14
Host controller
alert
Internal error
None
None
6
15
Host controller
alert
Internal error
None
None
6
16
Host controller
alert
Internal error
None
None
6
17
Host controller
alert
Internal error
None
None
6
18
Host controller
alert
Internal error
None
None