Troubleshooting, Lead-quality problems, Artifact” message on the screen – Welch Allyn CP50 - Quick Reference Guide User Manual

Page 13: Wandering baseline

Advertising
background image

Troubleshooting

Lead-quality problems

“Artifact” message on the screen

Artifact is signal distortion that makes it difficult to accurately discern the waveform

morphology.

Causes

The patient was moving.

The patient was shivering.

There is electrical interference.

Actions

See actions for wandering baseline, muscle tremor, and AC interference.

Wandering baseline

Wandering baseline is an upward and downward fluctuation of the waveforms.

Causes

Electrodes are dirty, corroded, loose, or positioned on bony areas.

The electrode gel is insufficient or dried.

The patient has oily skin or used body lotions.

Rising and falling of chest during rapid or apprehensive breathing.

Actions

Clean the patient’s skin with alcohol or acetone.

Reposition or replace the electrodes.

Verify that the patient is comfortable, warm, and relaxed.

If wandering baseline persists, turn the baseline filter on.

13