Welch Allyn Connex ProBP 3400, Understanding Battery Performance Capabilities - Quick Reference Guide User Manual

Connex, Probp, Understanding battery performance capabilities

Advertising
background image

Connex

®

ProBP

3400

Understanding Battery Performance Capabilities

To maximize ProBP battery 

performance ‐ and achieve 100 

readings per charge with a new 

battery ‐ the user must turn power 

off after each reading has been 

documented. 

If power is left on, the device will power 

down in 30 minutes*, however 

maximum cycles per charge will be 

significantly reduced.

Chart based on Average Use Case as described in table below

Battery Charge Cycles

~500

~ 100

Cy

cle

 Capacity

 (BP

 Re

ad

in

gs

)

of

 Fully

 Char

ge

d

 B

atte

ry

 †

FACTOR

Decreases Cycle Capacity 

of Fully Charged Battery

AVERAGE USE 

CASE

Increases Cycle Capacity
of Fully Charged Battery

MODE

N/A

STEP DEFLATION

SureBP Mode

INITIAL INFLATION

PRESSURE

(Step Deflation Mode Only)

Higher initial inflation pressure

160mmHg

Lower initial inflation pressure

CUFF SIZE

Larger blood pressure cuff

ADULT

Smaller blood pressure cuff

BLOOD PRESSURE

Higher Blood Pressure

120/80

Lower blood pressure

BATTERY AGE

Battery retains approximately 

75% of original capacity after 500 

charge cycles

NEW

N/A

† To fully charge ProBP battery, 

continue charging with power 

off for 1 hour after charge 

status indicator changes from 

amber to green.

Default settings:

2 Min. – Screen Blank

30 Min. – Power Off 

Adjustable using

Welch Allyn
Service Tool 

To turn power off after 

Screen Blank, press any 

button to reactivate display 

then press Power Button