Welch Allyn 40 inch Integrated Diagnostic System Wall Panel (76790-6) - Installation Guide User Manual

Installation

Advertising
background image

4341 State Street Road
Skaneateles Falls, NY 13153 USA
Phone: (800) 535-6663 (US Only) or (315) 685-4560 • Fax: (315) 685-3361
www.welchallyn.com

40” Integrated Wall Panel (order #76790-6)

Unpacking

Concealed damage: if you notice any damage after unpacking, notify the
transportation company at once to inspect the damage. Require them to give
you a “concealed” bad order report stating the condition of the goods when
examined. If necessary, insist it is their duty to perform this action.

If you comply with the above instructions, we will provide assistance in
establishing claims against the transportation company for loss and damage
in transit.

We do not accept goods returned for credit, exchange, or repairs, or for any
other reason, unless you have first contacted Welch Allyn and have our
written permission.

Installation

The Integrated Panel is mounted to the wall using a steel mounting bracket and
hardware (provided).

1.

Choose a location for the Integrated Panel near the examination table and an
electrical outlet. See Figure 1.

2.

Mark the location using the measurements shown in Figure 2.

3.

Locate and mark the studs in reference to the measurements in step 2.

For steps 4 through 6: when securing the mounting bracket to a wall with
metal studs, replace the drywall screws with sheet metal screws. When
securing the mounting bracket to a masonry/concrete block wall, replace the
drywall screws and wall anchors with masonry screws.

4.

Place the mounting bracket in the marked wall location and mark the
locations of all fasteners. Remove the mounting bracket and drill a 3/8” hole
for each wall anchor. Drill three holes if a wall stud is located or four holes if a
wall stud is not located.

5.

Install a wall anchor in each hole, if necessary. Install a screw in each wall
anchor and tighten enough to secure the wall anchor in place. Remove the
screws.

6.

Attach the mounting brackets to the wall with the screws. Level the mounting
bracket before securing it to the wall.

7.

Test the mounting bracket for strength.

8.

Mount the Welch Allyn products to the Integrated Panel using Figure 3, 4, or
5 to properly position the instruments.

Aneroid Blood Pressure Gauge: position the gauge bottom on the single pre-
installed screw and then secure the top of the gauge in place using the two
additional screws provided.

Hanging basket: position the thermometer on the two pre-installed screws,
and then secure in place using the two additional screws provided.

Install the remaining instrument assemblies on the pre-installed mounting
screws.

For Diagnostic Systems without SureTemp

®

thermometry, use the two

color-coated FastCaps. Remove the backing from the FastCap and adhere it
over the unused screw hole.

Figure 1. Suggested Room Layout

Manufactured for Welch Allyn by Heartwood. P/N: HRT 1037 Material No. 707809 Ver. C

Caution Welch Allyn is NOT responsible for the integrity of any wall
mounting interface. Welch Allyn recommends that the customer contact
their Biomedical Engineering Department or maintenance service to
ensure professional installation, safety, and reliability of any mounting
accessory.

CASEWORK

INTEGRATED

PANEL

STOOL

EXAM
TABLE

DRESSING NOOK

SIDE

CHAIR

10’-0”

8’-0”

Advertising
Table of contents