Welch Allyn Atlas Monitor Wall Mount Instructions - Installation Guide User Manual

Atlas, Wall mount installation guide, Using the ultimount™ wall mounting system

Advertising
background image

WARNING:

It is the health facility or other user’s responsibility to assure that the ULTIMOUNT™ wall mounting
system is anchored to a wall securely and consideration must be given to the load that will be placed
upon the wall mount. Each ULTIMOUNT™ is labeled with a maximum load rating of 33 pounds. Do
not exceed this load limit.

WARNING:

Welch Allyn, Inc. does not accept responsibility for any installation, nor for damage or injury arising from the
installation in any type of wall regardless of the type of fasteners and technique used.

WARNING:

It is the responsibility of the user of the ULTIMOUNT™ wall mount to see that it is securely anchored to the
wall and that the Atlas™ monitor is securely fastened to the wall mount as directed below.

Tools Required:

Phillips head Screw Driver

5/32” Hexagon Wrench (provided)

Hardware Included:

(1) Mounting Plate

(2) #8-32 Pan Head Phillips Screw - 3/8” long

(2) #8-32 Internal Tooth Lockwasher

(4) #10-32 Pan Head Phillips Screw - 1/2” long

(4) #10-32 Internal Tooth Lockwasher

Part No. 620426 Rev. A

Atlas

TM

Wall Mount Installation Guide

Using the ULTIMOUNT™ Wall Mounting System

Mounting Plate

Arm Assembly

Track

Basket

Locking Pin

Locking Pin

Track Stop Screws

Track
Hardware

Locking Pin Holes

Advertising