Welch Allyn 76790-1, Midmark Dressing Nook and 34 Integrated Panel With Diagnostic System - Installation Guide User Manual

Installation instructions

Advertising
background image

Installation

Instructions

Dressing Nook and 34" Integrated Panel
With Diagnostic System

Unpacking Notes

Concealed Damage. If there should be damage of
such nature that it could not be detected until the
goods were unpacked, have the Transportation
Company's Agent come AT ONCE and inspect them.
Require them to give you a "concealed" bad order
report stating the condition of the goods when exam-
ined. It is their duty to do this and you should INSIST
upon it.

If the Above Instructions are Complied With, we
shall cheerfully render all possible assistance in
establishing claims against the Transportation Com-
pany for loss and damage in transit.

We Do Not Accept Goods Returned For Credit,
Exchange, Repairs
, or for any other reason, unless
you have first communicated with us and secured our
written permission.

Installation

NOTE
The Dressing Nook and Integrated Panel are mounted to
the wall using steel mounting brackets (provided with
units). Wall anchors and drywall screws are also provided
to secure the mounting brackets to the wall. Figures 1
thru 5 are suggested mounting configurations and
locations only.

1.

Using the customer or room layout drawing, determine
where the Dressing Nook and / or Integrated Panel are
to be located. See Figure 1 for the suggested room
layout.

2.

Check clearance required for placing a side chair
beside the Dressing Nook. A minimum of 24 in. (61
cm) is required for the Midmark 429 or Ritter 129 Side
Chair. See Figure 2.

NOTE
If the locations of the units differ from what is shown in
Figure 2, then bracket locations shown in Figure 3 must be
changed accordingly.

3.

Measure and mark the mounting locations of the
mounting brackets on the wall. See Figure 3.

4.

Locate and mark the location of studs in reference to
mounting bracket; an electronic stud finder may be
useful.

NOTE
This note applies to steps 5 thru 7. When securing
mounting brackets to a wall with metal studs, use sheet
metal screws in place of drywall screws and when
securing mounting brackets to a masonry / concrete block
wall, use masonry screws in place of drywall screws and
wall anchors.

5.

Place the mounting bracket in position on wall and
mark the locations of all fasteners. Remove the
mounting bracket and drill a 3/8 in. hole where all wall
anchors are to be used; four holes per mounting
bracket if a wall stud is not located and three holes per
mounting bracket if a wall stud is located.

6.

Install a wall anchor in each hole. Install a screw in
each wall anchor and tighten far enough to secure the
wall anchor in place. Then, remove all screws from the
wall anchors.

EQUIPMENT ALERT
The 61-7/16 in. (156 cm) minimum
dimension between the Dressing Nook

mounting brackets is critical for installation.

7.

Attach mounting brackets to the wall using wall anchor
screws. Level each mounting bracket before securing
to the wall.

8.

Test each mounting bracket for strength.

9.

Mount the instruments to the Integrated Panel as
follows using Figure 4 or 5 to properly position the
instruments:
Position the Aneroid Blood Pressure Unit on the single
screw provided and then secure top of unit in place
using two additional screws provided. Install (hang)
the remaining instrument assemblies on the pre-
installed mounting screws.

10.

Lift the Dressing Nook and / or Integrated Panel onto
their mounting brackets.