Welch Allyn Connex Integrated Wall System (CIWS), Cardinal Health 34" Integrated Wall Panel - Installation Guide User Manual

Welch Allyn Equipment

Advertising
background image

1.

Choose a location for the panel near the examination table and an electrical
outlet.

Sample room layout

2.

On the selected wall, find and mark the studs, and choose the panel height
and corresponding height for the mounting rail bracket.

Recommendation: Place the bracket 63 in. (160 cm) from the floor.

3.

Affix the mounting rail bracket to two studs at the selected height using the
available screws (anchors are provided for additional support).

If studs are not available, use three anchors and screws.

4.

Hang the panel on the mounting rail bracket.

5.

Mount the Welch Allyn products to the panel according to the figures on the
reverse side of this instruction sheet.

Aneroid blood pressure gauge: position the gauge bottom on the pre-installed
screw and then secure the top of the gauge using the two additional screws
provided.

SureTemp

®

Plus thermometer: position the thermometer on the pre-installed

screws and then secure it with the two additional screws provided.

Install the remaining products on the pre-installed screws.

For diagnostic systems without SureTemp thermometry, cover the unused
screw holes with the FastCaps provided.

WARNING Welch Allyn nor Cardinal Health are responsible for the
integrity of any mounting installation. Welch Allyn recommends that
customers contact their Biomedical Engineering Department or
maintenance service to ensure professional installation for safety and
reliability of any mounting accessory.

34

 Integrated Wall Panel (

77790-CL)

Manual part no. 722465, 80019448 Ver. A
© 2012 Welch Allyn

1. Panel
2. Examination table

63 in

(160 cm)

12 in

(30 cm)

34 in

(86 cm)

Caution Ensure that the upper "lip" of the bracket sticks out from the
wall and that the bracket is level.

WARNING Ensure that the channel on the back of the panel fully
engages the mounting rail bracket. The panel should be level and flush
to the wall.