Best practices, Recommended network best practices, 2 - best practices – Welch Allyn Connex Central Station Motorola Best Practices-Network - Installation Guide User Manual

Page 7

Advertising
background image

2

Best practices

Recommended network best practices

The best practices and configuration settings listed in the following table are specific to
this vendor, and are recommended for the best performance. Increased data packet loss
or occasional disconnects are likely if these recommendations are not followed.

Best practice

Affected types Without best practice

Firmware version Use Welch-Allyn validated versions of the
Motorola Controller firmware

WiNG 5.4.1

Wireless

Using a non-validated version of firmware might cause
data loss.

Controller “Smart OCS Monitoring Frequency” — Change default
from 6s to 120s

Wireless

Loss of data (dropped UDP packets)

Hardware supported
Controllers

RFS6000

APs

AP300

AP6532

AP650*

Wireless

Using non-supported hardware might cause data loss.

* Tested for WPA-2 Personal use only, not validated for
WPA2 Enterprise.

Advertising