Nextteq Airflow Test Kit User Manual

Nextteq Accessories for air

Advertising
background image

OPERATING INSTRUCTIONS

P/N NSD0X0000003 / NSD0X0000004

Revision: 08/21/08 C

NEXTTEQ

LLC Page

1

of

2

8406 Benjamin Rd., Suite J
Tampa, FL 33634 USA
Toll free Tel: (877) 312-2333 / (813) 249-5888
Toll free Fax: (877) 312-2444 / (813) 249-0188
www.nextteq.com / [email protected]

AIRFLOW TEST KIT

Part No. 2106 (Entire Kit) / Part No. NST0X0002106 (Tubes Only)

Test Method Description from the OSHA Tuberculosis Control Training & Resource Guide:

1. Before using this product, read, understand, and comply with all labels, warnings, operating instructions, MSDS, and other

literature accompanying the product before use. Failure to comply with all labels, warnings, operating instructions, MSDS,
and other literature could result in serious injury.

2. Keep Nextteq LLC – MSDS P/N 90050 Rev. C, Dated August 21, 2008 for reference and training:
3. Wear eye protection and chemical resistant gloves in accordance with OSHA standards and good practice before using

this product.

4. Remove the aspirator bulb, black plastic tubing, and one plastic smoke tube from the Nextteq carrying case.
NOTE: To increase density of smoke, at temperatures colder than 40ºF/4.44ºC, we recommend that the tube be warmed with

the hand during use.

5. Connect the black plastic tubing to the aspirator bulb, then insert the black plastic tubing into the end of the smoke tube.
6. Break the two (2) ampoules in the plastic smoke tube by gently squeezing each glass ampoule.
NOTE: When the ampoules are manually crushed, aspirated airflow causes mixing of the released vapors, which react to

form Ethylenediamine acetate smoke.

7. The airflow tube should be held in the following position in relation to the door:

Parallel to the door (not aimed toward it)

Close to the floor

Centered in the middle of the door

Approximately two (2) inches in front of the door

NOTE: The tube should NOT be aimed toward the door and the smoke should NOT be blown directly into the isolation room.

8. Slowly squeeze the aspirator bulb pump to make small puffs of smoke, then observe the resulting direction of airflow.
NOTE: Minimize the momentum of the smoke by squeezing the bulb slowly, which will also help minimize the volume of

smoke released.

9. Repeat the test at least once or until consistent results are obtained.
10. For occupied isolation rooms, release smoke from the corridor side of the door only.

Smoke Tube Specifications:

Operating Temperature Range

40-90ºF (4.44-32.22ºC)

Operating Humidity Range

10-80% RH, non condensing

Storage Temperature Range

40-77ºF (4.44-25ºC)

Storage Humidity Range

10-80% RH, non condensing

Reuse of Tube

Plug ends of tube with rubber stoppers after first use if you plan on using the
tube again. Tubes need to be reused within 24 hours. Dispose of tubes after
24 hours if not reused.

Shelf Life

5 years from the date of shipment from Nextteq if stored correctly.

Required Storage Conditions

Store in a cool dry location out of direct sunlight.

Disposal

Seal ends of exhausted tubes and dispose in accordance with local, state, and
federal regulations.


Warnings

1.

Read, understand, and comply with all labels, warnings, operating instructions, MSDS, and other literature accompanying the product before use. Failure to
comply with all labels, warnings, operating instructions, MSDS, and other literature could result in serious injury.

2.

Operate the product in strict compliance with the manufacturer’s specifications and operating instructions.

3.

Be thoroughly familiar with this product, the manufacturer’s instructions, warnings, and associated equipment before using this product. Keep the Nextteq
LLC Material Safety Data Sheet, MSDS P/N 90050 Rev. 8 B, Dated August 5, 2008 for reference and training.

4.

Inspect the product and associated equipment for damage or defects before using this product. DO NOT use this product if this product or associated
equipment is damaged or defective.

5.

DO NOT remove, alter, or cover any label, instruction, or warning when using this product.

6.

Wear chemical resistant safety gloves in accordance with OSHA standards and good practice when using this product.

7.

Wear eye protection in accordance with OSHA standards and good practice when exposed to smoke or fumes.

8.

When operating the aspirator bulb pump, slowly and gently compress the bulb half way to minimize smoke momentum and smoke volume.

9.

Keep all bystanders away from this product during use or testing.

10. Smoke trail visualization isolation room testing should be performed in strict compliance with all local, state, and federal regulations.
11. DO NOT spray smoke or fumes of smoke tube under doors unless you have first checked to ensure the smoke will not be blown onto unprotected persons.