Calibration – US Balance US-PRISM 500g x 0.1g User Manual

Page 3

Advertising
background image

Pocket Scale

US-STINGER 500g X 0.1g

3


CALIBRATION


When to calibrate

Calibration may be required when it is initially installed, if the mini scale is moved to a substantial distance.
This is necessary because the weight a mass in one location is not necessarily the same in another location.
Also, with time and use, mechanical deviations may occur.

How to calibrate
Before entering calibration mode, press [ON/OFF] to turn the power on for 1 minute.

Enter calibration mode

Press [ON/OFF] to turn the power off.

Press [ZERO] first, then Press [ON/OFF],

While keeping pressing [ZERO], the display will show the A/D value.


Calibration by the weight

Wait for 3 seconds

Press [ON/OFF] the display will show “2Ero”, after 2 seconds, the display will show calibration weight

value.

Place the calibration weight on platform. Wait for 3 seconds, then press [ON/OFF],
The display will show “CAL”, after 2 seconds, the displayed will return to “PASS” and then return to A/D
value,

Now calibration is finished.

Press [ON/OFF], turn off the power,

Press [ON/OFF] turn on the power again,

Place the weight on the platform to make sure whether weighing is correct, if not, repeat

steps 2-5.

Return to weighing mode

Press [ON/OFF] to turn the power off.

Press [ON/OFF] again to turn on the

Power and the scale return to the weighing mode.