Dillon FI-521 User Manual
Page 31

FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
31
Response:
<LF>^^^^^^^^U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---over capacity
<LF>_ _ _ _ _ _ _ _ U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---under capacity
<LF> - - - - - - - - U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---zero-point error
Note:
U
1
U
2
U
3
is 1or 3 bytes according to current unit: kgf, lbf, N
<LF><P>W
1
W
2
W
3
W
4
W
5
<DP>W
6
U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---normal data
Note: The decimal point position is determined by CONFIG-PRIM.D
25.6.1.2 Command: S<CR> (53h 0dh) , request current status
Response: <LF> H
1
H
2
H
3
H
4
<CR><ETX>
25.6.1.3 Command: Z<CR> (5ah 0dh)
Response: Zero function is activated (simulate ZERO key) and it returns to current indicator status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
If ZERO function cannot be activated, it will return to current indicator status.
25.6.1.4 Command: U<CR> (55h 0dh)
Response: Changes units of measure (simulate UNIT key) and return indicator status with new units,
The new measure unit should be allowed to use
<LF> U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>
25.6.1.5 Command: L<CR> (4ch 0dh)
Response: Select next load cell to work (simulate CELL key), and returns indicator status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
25.6.1.6 Command: X<CR> (58h 0dh)
Response: power off the indicator, just like press down the ON/OFF key to turn off the indicator.
25.6.1.7 Command: all others
Response: Unrecognized command <LF>? <CR><ETX>
25.6.2 Summary of Command and Response:
Command
Response
ASCII HEX
W<CR> 57 0d
Read scale weight:
<LF>^^^^^^^^U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---over capacity
<LF>________U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---under capacity
<LF>-------- U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---zero-point error
<LF><p>W
1
W
2
W
3
W
4
W
5
<dp>W
6
U
1
U
2
U
3
<CR><LF>H
1
H
2
H
3
H
4
<CR><ETX>---norm
al data
S<CR> 53
0d
<LF>
H
1
H
2
H
3
H
4
<CR><ETX>; read indicator status
Z<CR> 5a
0d
<LF>
H
1
H
2
H
3
H
4
<CR><ETX> ; simulate ZERO key
U<CR> 55
0d
<LF>
U
1
U
2
U
3
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>; simulate UNIT key
L<CR>
4c 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX>; simulate SELECT key