Dillon FI-521 User Manual
Page 32

FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
32
X<CR>
58 0d power off the indicator, simulate OFF key
others
<LF>? <CR><ETX>
25.7 Communication when USER-COM1/2-LAYOUT is set to MULTIPLE:
25.7.1 Output string frame:
<LF><Add><Prompt><p>W
1
W
2
W
3
W
4
W
5
<Dp>W
6
U
1
U
2
U
3
<CR>
…… ---- Line number and content are determined by setting of USER-OUT1/2-xxxx
<LF><Add><Prompt>H
1
H
2
H
3
H
4
<CR> ---- USER-OUT1/2-STATUS is set to YES
……
<LF><Add><CR> --- USER-OUT1/2-LINE is set to LINE1…9
…… The number of blank lines is determined by USER-OUT1/2-LINE setting
<ETX> --- Last byte of string frame
25.7.2 Caution:
(1)The decimal point position is determined by CONFIG-PRIM.D
(2)The unit position and bytes is determined by which current unit is used.
(3)The details of <Prompt> refer to the content in 4.3USER Submenu.
(4)If USER-COM2-EN.ADDR is set to NO, no <Add> will be output.
(5)In continues output mode, if USER-COM1/2-LAYOUT is set to MULTIPLE, and many contents are
selected to output, the output contents from COM1 or COM2 may not catch up with the data processed
in indicator, So, if you want to watch “real time” data, you need to select fewer output contents and set
higher baud rate for COM1 or 2.
25.7.3 Examples of some layout when USER-OUT1-xxxx is set to YES:
IND.
ID:
123456
CELL No. : 1
CELL
TYPE:
COMPRESS
DSP.POL: POSITIVE
FORCE: 123.34 lbf
PEAK:
150.60 lbf
VALLEY:
10.78lbf
Max:
192.24lbf
Min:
1.56lbf
DATE: 2012-02-12
TIME: 12:34:56
A/D CODE: 982463
INPUT: 1.23034 mV/V
VOLTAGE: 6.7V
STATUS: bpq2