18 en message codes, Care, Warranty – Leica Geosystems Leica DISTO E7400x User Manual

Page 19: Safety instructions, Areas of responsibility

Advertising
background image

Leica DISTO

TM

E7400x 788480a

18

EN

Message Codes

If the message Error does not disappear
after switching on the device repeatedly,
contact the dealer.
If the message InFo appears with a number,
press the Clear button and observe the
following instructions:

Care

• Clean the device with a damp, soft cloth.
• Never immerse the device in water.
• Never use aggressive cleaning agents or

solvents.

Warranty

The device comes with a 3-year warranty. To
receive the 3-year warranty, the product
must be registered on www.disto.com
within 8 weeks of the purchase date. If the
product is not registered, a 2-year warranty
applies.

Safety Instructions

The person responsible for the instrument
must ensure that all users understand these
directions and adhere to them.

Areas of responsibility

Responsibilities of the manufacturer
of the original equipment:
Leica Geosystems AG
Heinrich-Wild-Strasse
CH-9435 Heerbrugg
Internet: www.disto.com
The company above is responsible for
supplying the product, including the User
Manual in a completely safe condition.
The company above is not responsible for
third party accessories.
Responsibilities of the person in
charge of the instrument:
• To understand the safety instructions on

the product and the instructions in the
User Manual.

• To be familiar with local safety regulations

relating to accident prevention.

• Always prevent access to the product by

unauthorised personnel.

No.

Cause

Correction

156

Transverse tilt greater
than 10°

Hold the instrument
without any transverse tilt.

162

Calibration mistake

Make sure, the device is
placed on a absolutely
horizontal and flat surface.
Repeat the calibration
procedure. If the mistake
still occurs, contact your
dealer.

204

Calculation error

Perform measurement
again.

252

Temperature too high Let device cool down.

253

Temperature too low Warm device up.

255

Received signal too
weak, measuring time
too long

Change target surface
(e.g. white paper).

256

Received signal too
high

Change target surface
(e.g. white paper).

257

Too much back-
ground light

Shadow target area.

258

Measurement outside
of measuring range

Correct range.

260

Laser beam inter-
rupted

Repeat measurement.

Advertising