Brookfield DV2+Pro Viscometer User Manual

Page 95

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 95

Manual No. M03-165-F0612

❏

Viscometer Will Not Return to Zero

✓

Viscometer is not level

• Check with spindle out of the sample

• Adjust the laboratory stand

✓

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

✓

Remove the spindle and turn the motor OFF; select display to % torque mode.

✓

Gently push up on the viscometer coupling.

✓

Turn the coupling until the digital display reads 10-15 on the % display.

✓

Gently let go of the coupling.

✓

Watch the digital display; you should see a “run” of numbers next to the %; the “run” of

numbers should ultimately stop at 0.0 (+/- 0.1).

If the digital display does not return to ZERO, the unit most likely is in need of service.

• Perform calibration check (See Appendix E)

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair (See

Appendix N).

* This procedure does not apply to instruments with ball bearing suspension. (See section I.3)

❏

Display Reading Will Not Stabilize

✓

Special characteristic of sample fluid. There is no problem with the viscometer.

• Refer to Appendix C

✓

Check for erratic spindle rotation.

• Verify power supply

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

✓

Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

✓

Temperature fluctuation in sample fluid.

• Use temperature bath for control

❏

No Recorder Response

✓

Be Sure the viscometer is not at ZERO reading.

✓

Be sure the recorder is ON and not on STANDBY.

✓

Verify the range settings.

✓

Check cable leads for clean connection.

✓

Verify cable connections (see Appendix H).