Manual No. M13-167-B0614

❏

Viscometer Will Not Return to Zero

✓

Viscometer is not level

• Check with spindle out of the sample.

• Adjust the laboratory stand.

✓

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

✓

Remove the spindle and turn the motor OFF; select display to % torque mode.

✓

Gently push up on the viscometer coupling.

✓

Turn the coupling until the digital display reads 10-15 on the % display.

✓

Gently let go of the coupling.

✓

Watch the digital display; you should see a “run” of numbers next to the %; the

“run” of numbers

should ultimately stop at 0.0 (+/- 0.1).

If the digital display does not return to ZERO, the unit most likely is in need of

service.

• Perform calibration check (see Appendix F).

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair (see Appendix O).

* This procedure does not apply to instruments with ball bearing suspension (see Section

I.3).

❏

Display Reading Will Not Stabilize

✓

Special characteristic of sample fluid. There is no problem with the viscometer.

• Refer to Appendix C.

✓

Check for erratic spindle rotation.

• Verify power supply

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair.

✓ Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for

repair.

✓

Temperature fluctuation in sample fluid.

• Use temperature bath for control.