The full scale viscosity range for any DV-II+ model and spindle may be calculated using the
equation:
Full Scale Viscosity Range [cP] = TK * SMC *
10,000
RPM
where:
TK
= DV-II+ Torque Constant from Table D2
SMC = Spindle Multiplier Constant from Table D1
The Shear Rate calculation is:
Shear Rate (1/Sec) = SRC * RPM
where:
SRC = Shear Rate Constant from Table D1