Brookfield CT3 Texture Analyzer User Manual
Page 7

Brookfield Engineering Labs., Inc.
Page 7
Manual No. M08-372-C0113
Trigger Point:
Range
Resolution
Model
Grams
Grams
CT3-100
0.1-10
0.01
CT3-1000
0.2-100 0.10
CT3-1500
0.2-150 0.2
CT3-4500
0.5-500 0.5
CT3-10KG
1-1000
1.0
CT3-25KG
2-2500
2.0
CT3-50KG
5-5000
5.0
Recommended trigger value settings are given in Table II.1
NOTE: Setting a trigger value of zero disables trigger function so test will immediately start
from current position.
Speed: 0.01 to 0.1 mm/s in increments of 0.01 mm/s
0.1 to 10 mm/s in increments of 0.1 mm/s
Accuracy: ±0.1% of set speed
Position:
Range: 0-101.6 mm
Resolution: 0.1mm
Accuracy: 0.1mm
Increments for changing test speed using the Select/Scroll knob can be switched between
0.1mm/s and 1.0mm/s by depressing Reset/Stop button.
Temperature:
-20°C to 120°C (Requires optional probe, DVP-94Y)
Output:
RS232 Compatible Serial Port, USB Port
Environmental
0°C to 40°C temperature range (41°F to 104°F)
Conditions:
20% - 80% relative humidity, non-condensing atmosphere
Use:
Intended for indoor use only
Altitude: up to 2000m
Dimensions:
10.5” x 10.5” x 24”
Weight:
37 lbs. (16.8 kg)
I.5 Installation
1) Prepare a clean, level surface.
NOTE: This instrument is a sensitive force measuring device. It should be installed on a
clean, solid, level bench surface which is free from external vibrations.
2) Unpack and remove the CT3 Texture Analyzer from the shipping container.
The CT3 Texture Analyzer weighs 16.8 kg (37 pounds). Use caution when lifting the
unit out of the packaging.
3) Place the CT3 Texture Analyzer on a sturdy, level surface. Adjust the four feet to ensure
that the instrument is stable.